f y in p b t

+8615280858852

sales@avenotester.com

Englishen españoles portuguêspt

Technical

AC04 AATCC Electronic Crockmeter Maintenance

2019-07-26 16:42:26

1. Friction head for plastic products, have been finished friction before going out factory, should no gap with stainless steel plate, friction head should be touched directly with sandpaper when uses. It is easy friction head to wear rather than flat. Larger gap with steel plate results in uneve pressure from the samples and impacting on testing results.

2. The speed of friction head has been adjusted, it is inadvisable to rotate adjustment knob of regulator.

3. Friction driving end arm (in-house equipment) have applied lubricant to ensure smoothly move and reduce noise, if AATCC Electronic Crockmeter friction noise of the apparatus largens in testing process, could be appropriate to add lubricant.

4. Show wear with sandpaper to test in the test process, so should exchange sandpaper regularly.

5. When start-up AATCC Crock Meter, it doesn’t work, so should check regulator whether is in working condition (pressing an end of white point) or check the fuse whether breaks up.


AVENO Machine Recommend:

AC04 AATCC Electronic Crockmeter

Contact Us Now!

Sales Dept Tel: +86 15280858852

Email: sales@avenotester.com

Skype: sales@avenotester.com

Web: www.avenotester.com



 
chat now please click here for inquiry
Any questions we will reply you within in 24 hours. Best wishes.