f y in p b t

+8615280858852

sales@avenotester.com

Englishen españoles portuguêspt

Technical

Calibration procedure Of Safety shoes Inserts Flexing Tester

2021-11-10 10:58:13

1. Correction shoes Inserts Flexing Tester items: folding distance and speed.

2. Calibration equipment: height gauge, infrared speed measuring device

3. Calibration cycle: one year

4. Correction steps:

a) Verification of endurance distance

i. First manually adjust the clamp to the lowest point, use a height ruler to measure the height of the rubber interlayer of the two fixed rods (middle), and record.

ii. Manually rotate the knob to turn the fixture. When it reaches the highest position, use a height ruler to measure the height of the rubber interlayer between the two fixed rods (middle) and record it.

iii. Calculate the difference between the two recorded values and whether it is within the range compared with the standard value.

b) Speed correction method:

i. Start the POWER button, reset the value in the "PV" column on the control panel to zero, and set the number of tests in the "SV" column.

ii. Press the START button to start the shoes Inserts Flexing Testing machie.

iii. Use an infrared speedometer to measure the number of revolutions of the shaft and compare the difference with the standard value.


More about Safety shoes Inserts Flexing Tester please click:

https://www.avenotester.com/safety-shoes-inserts-flexing-tester-af06_p62.html

shoes Inserts Flexing Testing machie


Contact Us Now!

Sales Dept Tel: +86 15280858852

Email: sales@avenotester.com

Skype: sales@avenotester.com

Web: www.avenotester.com



 
chat now please click here for inquiry
Any questions we will reply you within in 24 hours. Best wishes.